Cardiol Res
Cardiology Research, ISSN 1923-2829 print, 1923-2837 online, Open Access
Article copyright, the authors; Journal compilation copyright, Cardiol Res and Elmer Press Inc
Journal website http://www.cardiologyres.org

Original Article

Volume 5, Number 3-4, August 2014, pages 101-107


Evaluation of a New Shirt-Based Electrocardiogram Device for Cardiac Screening in Soccer Players: Comparative Study With Treadmill Ergospirometry

Oscar Fabregat-Andresa, b, d, Adolfo Munoz-Machob, Guillermo Adell-Beltranb, Xavier Ibanez-Catalac, Agustin Maciac, Lorenzo Facilaa

aDepartment of Cardiology, Consorcio Hospital General Universitario, Valencia, Spain
bVillarreal Club de Futbol SAD Medical Services, Villarreal, Spain
cNuubo, Madrid, Spain
dCorresponding Author: Oscar Fabregat-Andres, Department of Cardiology, Consorcio Hospital General Universitario de Valencia and Villarreal Club de Futbol SAD Medical Services, Villarreal, Spain

Manuscript accepted for publication April 25, 2014
Short title: Electrocardiogram Device for Cardiac Screening
doi: https://doi.org/10.14740/cr333w

Abstract▴Top 

Background: Prevention of cardiac events during competitive sports is fundamental. New technologies with remote monitoring systems integrated into clothing could facilitate the screening of heart disease. Our aim was to evaluate the feasibility of Nuubo system during a field stress test performed by soccer players, comparing results with treadmill ergospirometry as test reference.

Methods: Nineteen male professional soccer players (19.2 ± 1.6 years) were studied. Wireless electrocardiographic monitoring during a Yo-Yo intermittent recovery test level 1 in soccer field and subsequent analysis of arrhythmias were firstly performed. Subsequently, in a period no longer than 4 weeks, each player underwent cardiopulmonary exercise testing in hospital.

Results: During Yo-Yo test, electrocardiogram (ECG) signal was interpretable in 16 players (84.2%). In the other three players, ECG artifacts did not allow a proper analysis. Estimation of maximum oxygen consumption was comparable between two exercise tests (VO2 max 53.3 ± 2.4 vs. 53.7 ± 3.0 mL/kg/min for Yo-Yo test and ergometry respectively; intra-class correlation coefficient 0.84 (0.63 - 0.93), P < 0.001). No arrhythmias were detected in any player during both tests.

Conclusions: The use of Nuubo’s technology allows an accurate single-lead electrocardiographic recording and estimation of reliable performance variables during exercise testing in field, and provides a new perspective to cardiac remote monitoring in collective sports.

Keywords: Cardiac screening; Wireless device; ECG monitoring; Exercise test; Ergospirometry; Soccer

Introduction▴Top 

Primary prevention of cardiac events in competitive sports has emerged as a key target of sports medicine. Screening protocol for cardiovascular abnormalities proposed by current guidelines [1, 2] consists at least of complete medical history, physical examination and resting 12-lead electrocardiogram (ECG). This screening strategy has provided solid evidence that may prevent sudden cardiac death in the athlete [3], although its large-scale implementation remains controversial [4].

In fact, due to the dynamic nature of ECG abnormalities in some channelopathies responsible for a high percentage of cases of sudden death in athletes, usefulness of resting ECG could be considered limited [5]. Thereby, some conditions such as increased body temperature, changes in autonomic status or increased heart rate are necessary for definitive diagnosis [6-8]. All these variables come together during intense physical exercise practice.

Ergospirometry as stress test is widely used in elite athletes. However, the accuracy of continuous and progressive tests to determine individual physical performance in team sports as soccer, which requires repeated anaerobic efforts with short resting times, remains still controverted, because the activity developed with treadmill test is suboptimally adjusted to the sport real practice.

For these reasons, new technologies with remote wireless monitoring systems integrated into clothing could facilitate the screening of heart disease in collective sports [9]. Secondly, these emergent technological tools could also provide a more accurate assessment of physical performance of the athlete as they allow the practice of specific exercise tests adapted to the sports field.

Nuubo’s dynamic ECG (nECG platform) is a new single-lead dynamic ECG wireless monitoring system that incorporates biomedical e-textile technology, and consists of three elements: an electronic device (nECG minder) attached to the garment that transmits the ECG signal via Bluetooth® to a computer (among other signals like accelerometer); a biomedical shirt that captures the ECG signal via textile electrodes technology BlendFix® (nECG shirt); and a software package that manages sessions, activities and users, and allows the visualization and analysis of data captured by the nECG minder (nECG suite) [10].

Our objective was to evaluate the applicability of Nuubo system during a stress test performed in field by professional soccer players, assessing both the ability of this device to screen cardiac arrhythmias during exercise and its capability to estimate individual physical performance using specific tests, compared with treadmill ergospirometry and test reference.

Material and Methods▴Top 

Participants

Nineteen male professional soccer players from Villarreal B team of the Spanish second division B were studied. Although the team consists of 23 players, four were excluded because of injury. The study protocol was consistent with the institutional ethical requirements in accordance with the Declaration of Helsinki and all subjects were fully informed about the procedures to be used.

Field test with Nuubo system

Yo-Yo intermittent recovery test (YYIRT) level 1 was performed at the beginning of training session after 10 min of warming. It was used as reference test to determine performance and indirect estimation of maximum oxygen consumption (VO2 max) in soccer players [11, 12]. An initial session was completed for 17 players. Other two injured players at the moment of first test, performed YYIRT 2 weeks later. All players wore an nECG shirt 15 min before exercise and the quality of the signal was checked by wireless transmission to a computer located in the field. Subsequently, ECG signal was analyzed using nECG suite software to detect arrhythmias and to determine parameters of heart rate variability (HRV). VO2 max was calculated using the formula: (distance in meters × 0.0084 + 36.4) [13].

Ergospirometry

All players performed a treadmill exercise testing in hospital within a month after the YYIRT. A progressive, continuous, triangular, maximum protocol was used, consisting of warming at 5 km/h for 5 min, exercise at 8 km/h for 2 min with increases of 2 km/h each 2 min, and recuperation at 5 km/h for 5 min. Expired gases were analyzed in each ventilatory cycle and VO2 max was directly calculated.

Other complementary tests

Before training, fasting blood test, 12-lead ECG at rest and blood pressure measurement were performed. Peripheral lactate was also analyzed before and immediately after the test using a Lactate Pro 2 Analyzer (Arkray Inc., Japan).

Statistical analysis

Quantitative data were expressed as mean ± standard deviation and qualitative data as median and absolute range. Normal distribution was analyzed by Kolmogorov-Smirnov test. Comparisons between groups were made with t-test for continuous variables. Inter-methods agreement was evaluated by means of the intra-class correlation coefficient (ICC) and kappa index, and the relationship between both tests was also calculated using Pearson’s correlation coefficient (r). Differences were considered statistically significant when a P value < 0.05 was obtained. All analyses were performed using SPSS version 17.0 (SPPS Inc., USA).

Results▴Top 

Nineteen players completed both exercise tests. Baseline characteristics of the participants, blood pressure measurement, lactate values, YYIRT and ergospirometry results, and HRV parameters estimated by Nuubo system are shown in Table 1.

Table 1.
Click to view
Table 1. Baseline Characteristics of Participants, Yo-Yo Test and Ergospirometry Results, and Heart Rate Variability Parameters Estimated by Nuubo System
 

Both tests showed an appropriate inter-methods agreement, as shown in Table 2. Analyzing feasibility of Nuubo system in sports field, we found that ECG signal was adequate and interpretable in 16 players (84.2%) both at rest and during the three phases of the exercise: warming up, YYIRT and recovery time, clearly differentiated in activity index graph (Fig. 1). All these individuals presented mild signal abnormalities affecting less than three consecutive QRS complexes that did not prevent a proper interpretation of ECG, inasmuch as nECG suite software analysis was able to differentiate adequate QRS complexes from those considered artifact.

Table 2.
Click to view
Table 2. Comparative Results Between Yo-Yo Test With Nuubo’s Technology and Ergospirometry
 

Figure 1.
Click for large image
Figure 1. Wireless electrocardiographic monitoring recordings of soccer players during Yo-Yo intermittent recovery test level 1. (A) Activity index during the three phases of the test: warming up, Yo-Yo test and recovery. Physical activity expressed in units of acceleration (g). (B) ECG recording of a patient with multiple artifacts in baseline signal, impeding an accurate analysis of QRS complexes. (C) Interpretable ECG recording.

When analyzing cardiac abnormalities, no ventricular extrasystoles or other arrhythmias were detected during any tests. Nor significant ST depression was detected in any player during both methods.

In order to assess the usefulness of Nuubo system during YYIRT to determine physical performance variables compared with ergospirometry, we found a good correlation in calculation of VO2 max (VO2 max 53.3 ± 2.4 vs. 53.7 ± 3.0 mL/kg/min for Yo-Yo test and ergometry respectively; intra-class correlation coefficient 0.84 (0.63 - 0.93), P < 0.001) (Fig. 2). Maximal heart rate and maximal lactate reached by players are similar in both tests (comparisons using t-test: 187.7 ± 3.8 vs. 190.1 ± 5.5 beats per minute, P = 0.8; 10.4 ± 1.7 vs. 8.5 ± 3.3 mmol/L, P = 0.3; for Yo-Yo test and ergometry respectively). In addition, ergospirometry allowed the calculation of aerobic and anaerobic thresholds adequately (28.5 ± 6.1 and 38.5 ± 4.1 mL/kg/min respectively).

Figure 2.
Click for large image
Figure 2. Relationship between treadmill ergospirometry and Yo-Yo intermittent recovery test level 1 in the calculation of maximum oxygen consumption (VO2 max).
Discussion▴Top 

Screening of cardiac rhythm and conduction disturbances is universally supported to identify athletes with pre-existing conditions that place them at risk of catastrophic injury or sudden death. The pre-participation assessment as recommended by current guidelines involves a focused player and family medical history, a cardiac specific physical examination and resting 12-lead ECG [14].

Although other diagnostic tests are not initially required, the prognostic value of exercise testing in competitive athletes is undoubted. At present, although ergospirometry is widely used by elite teams because it allows direct estimation of performance variables such as VO2 max or aerobic/anaerobic thresholds, unfortunately it has the disadvantage of the need to travel to a hospital with consequent loss of training sessions. Furthermore, treadmill test is a continuous and progressive test that is poorly adapted to efforts of soccer players during trainings and matches. Thus, the implementation of medical tests is imperative in sports field that would allow continuous and reliable cardiac monitoring of soccer players since they represent an unsolved problem.

Accordingly, the use of electronic devices adapted to clothing has been recently evaluated in outpatients as those in cardiac rehabilitation [15]. Nonetheless, those have been poorly studied in sport as yet [16]. In this pilot study, Nuubo demonstrates for the first time its applicability in monitoring soccer players during a stress test in field, and seems to be a suitable system to reliably detect arrhythmias and calculate HRV parameters. Further studies will therefore be needed to implement this technology in the context described.

The main limitation of the study is the small sample during a single test. However, since it is a novel technology, this work could serve to spread this system in screening heart disease and assess the physical performance of soccer teams.

Conclusions

The use of Nuubo’s technology allows an accurate single-lead electrocardiographic recording and estimation of reliable performance variables during exercise testing in field, and provides a new perspective to cardiac remote monitoring in team sports as soccer.

Grant Support

None.

Financial Disclosures

Xavier Ibanez-Catala and Agustin Macia are Nuubo employees.


References▴Top 
  1. Corrado D, Pelliccia A, Bjornstad HH, Vanhees L, Biffi A, Borjesson M, Panhuyzen-Goedkoop N, et al. Cardiovascular pre-participation screening of young competitive athletes for prevention of sudden death: proposal for a common European protocol. Consensus Statement of the Study Group of Sport Cardiology of the Working Group of Cardiac Rehabilitation and Exercise Physiology and the Working Group of Myocardial and Pericardial Diseases of the European Society of Cardiology. Eur Heart J. 2005;26(5):516-524.
    doi pubmed
  2. Maron BJ, Thompson PD, Ackerman MJ, Balady G, Berger S, Cohen D, Dimeff R, et al. Recommendations and considerations related to preparticipation screening for cardiovascular abnormalities in competitive athletes: 2007 update: a scientific statement from the American Heart Association Council on Nutrition, Physical Activity, and Metabolism: endorsed by the American College of Cardiology Foundation. Circulation. 2007;115(12):1643-1455.
    doi pubmed
  3. Corrado D, Basso C, Pavei A, Michieli P, Schiavon M, Thiene G. Trends in sudden cardiovascular death in young competitive athletes after implementation of a preparticipation screening program. JAMA. 2006;296(13):1593-1601.
    doi pubmed
  4. Sharma S, Estes NA 3rd, Vetter VL, Corrado D. Clinical decisions. Cardiac screening before participation in sports. N Engl J Med. 2013;369(21):2049-2053.
    doi pubmed
  5. Martin CA, Matthews GD, Huang CL. Sudden cardiac death and inherited channelopathy: the basic electrophysiology of the myocyte and myocardium in ion channel disease. Heart. 2012;98(7):536-543.
    doi pubmed
  6. Morita H, Zipes DP, Morita ST, Wu J. Temperature modulation of ventricular arrhythmogenicity in a canine tissue model of Brugada syndrome. Heart Rhythm. 2007;4(2):188-197.
    doi pubmed
  7. Fossa AA. The impact of varying autonomic states on the dynamic beat-to-beat QT-RR and QT-TQ interval relationships. Br J Pharmacol. 2008;154(7):1508-1515.
    doi pubmed
  8. Cherry EM, Fenton FH, Gilmour RF Jr. Mechanisms of ventricular arrhythmias: a dynamical systems-based perspective. Am J Physiol Heart Circ Physiol. 2012;302(12):H2451-2463.
    doi pubmed
  9. Despang HG, Netz S, Heinig A, Holland HJ, Fischer WJ. Wireless long-term ECG integrated into clothing. Biomed Tech (Berl). 2008;53(6):270-278.
    doi pubmed
  10. Perez de Isla L, Lennie V, Quezada M, Guinea J, Arce C, Abad P, Saltijeral A, et al. New generation dynamic, wireless and remote cardiac monitorization platform: a feasibility study. Int J Cardiol. 2011;153(1):83-85.
    doi pubmed
  11. Rampinini E, Sassi A, Azzalin A, Castagna C, Menaspa P, Carlomagno D, Impellizzeri FM. Physiological determinants of Yo-Yo intermittent recovery tests in male soccer players. Eur J Appl Physiol. 2010;108(2):401-409.
    doi pubmed
  12. Karakoc B, Akalan C, Alemdaroglu U, Arslan E. The relationship between the yo-yo tests, anaerobic performance and aerobic performance in young soccer players. J Hum Kinet. 2012;35:81-88.
    doi pubmed
  13. Bangsbo J, Iaia FM, Krustrup P. The Yo-Yo intermittent recovery test : a useful tool for evaluation of physical performance in intermittent sports. Sports Med. 2008;38(1):37-51.
    doi pubmed
  14. Dvorak J, Kramer EB, Schmied CM, Drezner JA, Zideman D, Patricios J, Correia L, et al. The FIFA medical emergency bag and FIFA 11 steps to prevent sudden cardiac death: setting a global standard and promoting consistent football field emergency care. Br J Sports Med. 2013;47(18):1199-1202.
    doi pubmed
  15. Skobel E, Martinez-Romero A, Scheibe B, Schauerte P, Marx N, Luprano J, Knackstedt C. Evaluation of a newly designed shirt-based ECG and breathing sensor for home-based training as part of cardiac rehabilitation for coronary artery disease. Eur J Prev Cardiol. 2013;
    pubmed
  16. Perego P, Moltani A, Andreoni G. Sport monitoring with smart wearable system. Stud Health Technol Inform. 2012;177:224-228.
    pubmed


This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.


Cardiology Research is published by Elmer Press Inc.

 

Browse  Journals  

 

Journal of Clinical Medicine Research

Journal of Endocrinology and Metabolism

Journal of Clinical Gynecology and Obstetrics

 

World Journal of Oncology

Gastroenterology Research

Journal of Hematology

 

Journal of Medical Cases

Journal of Current Surgery

Clinical Infection and Immunity

 

Cardiology Research

World Journal of Nephrology and Urology

Cellular and Molecular Medicine Research

 

Journal of Neurology Research

International Journal of Clinical Pediatrics

 

 
       
 

Cardiology Research, bimonthly, ISSN 1923-2829 (print), 1923-2837 (online), published by Elmer Press Inc.                     
The content of this site is intended for health care professionals.

This is an open-access journal distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License, which permits unrestricted
non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Creative Commons Attribution license (Attribution-NonCommercial 4.0 International CC-BY-NC 4.0)


This journal follows the International Committee of Medical Journal Editors (ICMJE) recommendations for manuscripts submitted to biomedical journals,
the Committee on Publication Ethics (COPE) guidelines, and the Principles of Transparency and Best Practice in Scholarly Publishing.

website: www.cardiologyres.org   editorial contact: editor@cardiologyres.org
Address: 9225 Leslie Street, Suite 201, Richmond Hill, Ontario, L4B 3H6, Canada

© Elmer Press Inc. All Rights Reserved.


Disclaimer: The views and opinions expressed in the published articles are those of the authors and do not necessarily reflect the views or opinions of the editors and Elmer Press Inc. This website is provided for medical research and informational purposes only and does not constitute any medical advice or professional services. The information provided in this journal should not be used for diagnosis and treatment, those seeking medical advice should always consult with a licensed physician.